E-ISSN 2218-6050 | ISSN 2226-4485
 

Original Article


Comparison between the Parks Medical Doppler and the Mano Médical Vet BP Doppler for recording indirect systemic blood pressure in conscious dogs

Marina Domingues, Panagiotis Archontakis, Jonathan Bouvard, Luke Winter, Danielle Gunn-Moore.


Abstract
Objectives: To evaluate agreement between Parks Medical Doppler (PMD) and Mano Médical Vet BP Doppler (MMVBP) for assessing systemic blood pressure (SBP) in conscious dogs.
Methods: 40 client-owned dogs were prospectively enrolled; SBP measurements were acquired by a single operator using the PMD then the MMVBP. The mean of 5 consecutive measurements for each device was classified according to target organ damage (TOD) risk scores (1 < 140 mmHg; 2 =140-160 mmHg; 3 =160-180 mmHg; 4 ≥180 mmHg).
Results: Total mean SBP for the devices was not statistically different (p =0.77). However, Bland-Altman analysis revealed wide limits of agreement, with MMVBP slightly underestimating SBP compared to PMD (bias = –0.6mmHg, 95% LoA: – 26.3 – 25.09). Both devices correlated well (r = 0.8269; p < 0.0001) and had identically acceptable intra-observer repeatability (coefficients of variation of 4.09% for MMVBP; 3.86% for PMD). Four dogs (10%) had a TOD score of 3 by one device but scored < 3 with the other.
Conclusions: A good agreement and correlation was observed between the PMD and the MMVBP suggesting that both devices can be used interchangeably for assessment of SBP in conscious dogs. The wide limits of agreement observed between both devices were most likely associated with intraindividual variability in SBP over time.

Key words: Conscious, Dogs, Doppler, Systemic blood pressure


 
ARTICLE TOOLS
Abstract
PDF Fulltext
HTML Fulltext
How to cite this articleHow to cite this article
Citation Tools
Related Records
 Articles by Marina Domingues
Articles by Panagiotis Archontakis
Articles by Jonathan Bouvard
Articles by Luke Winter
Articles by Danielle Gunn-Moore
on Google
on Google Scholar


How to Cite this Article
Pubmed Style

Domingues M, Archontakis P, Bouvard J, Winter L, Gunn-Moore D, . Comparison between the Parks Medical Doppler and the Mano Médical Vet BP Doppler for recording indirect systemic blood pressure in conscious dogs. Open Vet J. 2022; 12(4): 562-566. doi:10.5455/OVJ.2022.v12.i4.19


Web Style

Domingues M, Archontakis P, Bouvard J, Winter L, Gunn-Moore D, . Comparison between the Parks Medical Doppler and the Mano Médical Vet BP Doppler for recording indirect systemic blood pressure in conscious dogs. https://www.openveterinaryjournal.com/?mno=103729 [Access: April 10, 2024]. doi:10.5455/OVJ.2022.v12.i4.19


AMA (American Medical Association) Style

Domingues M, Archontakis P, Bouvard J, Winter L, Gunn-Moore D, . Comparison between the Parks Medical Doppler and the Mano Médical Vet BP Doppler for recording indirect systemic blood pressure in conscious dogs. Open Vet J. 2022; 12(4): 562-566. doi:10.5455/OVJ.2022.v12.i4.19



Vancouver/ICMJE Style

Domingues M, Archontakis P, Bouvard J, Winter L, Gunn-Moore D, . Comparison between the Parks Medical Doppler and the Mano Médical Vet BP Doppler for recording indirect systemic blood pressure in conscious dogs. Open Vet J. (2022), [cited April 10, 2024]; 12(4): 562-566. doi:10.5455/OVJ.2022.v12.i4.19



Harvard Style

Domingues, M., Archontakis, P., Bouvard, J., Winter, L., Gunn-Moore, D. & (2022) Comparison between the Parks Medical Doppler and the Mano Médical Vet BP Doppler for recording indirect systemic blood pressure in conscious dogs. Open Vet J, 12 (4), 562-566. doi:10.5455/OVJ.2022.v12.i4.19



Turabian Style

Domingues, Marina, Panagiotis Archontakis, Jonathan Bouvard, Luke Winter, Danielle Gunn-Moore, and . 2022. Comparison between the Parks Medical Doppler and the Mano Médical Vet BP Doppler for recording indirect systemic blood pressure in conscious dogs. Open Veterinary Journal, 12 (4), 562-566. doi:10.5455/OVJ.2022.v12.i4.19



Chicago Style

Domingues, Marina, Panagiotis Archontakis, Jonathan Bouvard, Luke Winter, Danielle Gunn-Moore, and . "Comparison between the Parks Medical Doppler and the Mano Médical Vet BP Doppler for recording indirect systemic blood pressure in conscious dogs." Open Veterinary Journal 12 (2022), 562-566. doi:10.5455/OVJ.2022.v12.i4.19



MLA (The Modern Language Association) Style

Domingues, Marina, Panagiotis Archontakis, Jonathan Bouvard, Luke Winter, Danielle Gunn-Moore, and . "Comparison between the Parks Medical Doppler and the Mano Médical Vet BP Doppler for recording indirect systemic blood pressure in conscious dogs." Open Veterinary Journal 12.4 (2022), 562-566. Print. doi:10.5455/OVJ.2022.v12.i4.19



APA (American Psychological Association) Style

Domingues, M., Archontakis, P., Bouvard, J., Winter, L., Gunn-Moore, D. & (2022) Comparison between the Parks Medical Doppler and the Mano Médical Vet BP Doppler for recording indirect systemic blood pressure in conscious dogs. Open Veterinary Journal, 12 (4), 562-566. doi:10.5455/OVJ.2022.v12.i4.19